Skip to main content

Table 2 MRI sequences and parameters used in the study; S and T define Sagittal and Transversal orientation, respectively

From: Study protocol: the Whitehall II imaging sub-study

 

Structural

Functional

Sequence

MEMPR

FLAIR

T2*

dMRI

Multiband

Standard

Condition

————————————

————————————

———————————

————————————

Resting

Resting

TR in ms

2530

9000

36

8900

1300

3000

TE in ms

1.79/3.65/5.51/7.37

73

30

91.2

40

30

Flip angle

7°

150°

15°

————————————

66°

90°

Voxel in mm 3

1x1x1

0.9x0.9x3

0.7x0.7x1.5

2x2x2

2x2x2

3x3x3

FoV read

256

220

220

192

212

192

FoV phase

100%

100%

81.3%

100%

100%

100%

Base resolution

256

256

320

96

106

64

Phase resolution

100%

100%

100%

100%

100%

100%

TI in ms

1380

2500

———————————

————————————

————————————

————————————

Bandwidth

651 Hz/Px

283 Hz/Px

170 Hz/Px

1680 Hz/Px

1814 Hz/Px

2368 Hz/Px

Orientation

S

T

T

T

T

T

b-value

—————————

———————————

——————————

1500 s/mm2

———————————

—————————————

N. of volumes

—————————

———————————

—————————

——————————

460

200

N. of directions

——————————

———————————

——————————

60 + 5 b = 0 s

————————————

—————————————

Acquisition time

6 m 12 s

4 m 14 s

4 m 17 s

9 m 56 s

10 m 10 s

10 m